DataSim 6100

医療機器製造・販売の大正医科器械 TOP >> フルークバイオメディカル解析装置(MEテスタ) >> DataSim 6100

模擬波形発生装置 DataSim 6100

模擬波形発生装置 DataSim 6100

特長

  • Extensive training capability
  • Interface with Resusci-Anne, Arrhythmia Anne, and Chris Clean defibrillator mannequins
  • Synchronization of hemodynamic waveforms
  • Manual PAC and PVC insertions
  • Swan-Ganz procedure
  • Video adapter interface
  • 6 channels for generating ECG, arrhythmia, blood pressure and respiration
Power
Battery Type: 12 V, 1.9 AH sealed lead acid rechargeable
Battery Capacity: ten hours
Battery Charge Time: 5% to 95% of complete charge in 14 hours
ECG
Output signals
- High Level: 1 V/1 mV
Heart rate
- Range: 30 BPM to 300 BPM
- Accuracy: ± 1%
Output connectors
- Low Level: 12-Lead electrode snaps
- High Level: Switchcraft 15GM7F
- Pacemaker Artifact: -8 mV, 1 ms
- Sample Rate: 250 sample/s max
Performance testing
- Linearity: 2.5 Hz triangular wave
- HR Cal Check: 30 BPM to 300 BPM
- Chart Speed: 2.5 Hz square wave
- Amplitude: 0.25 mV, 0.5 mV, 1 mV, 1.5 mV, 2 mV, and 2.50 mV
Blood Pressure
Output signals
- High Level: 1V/100 mmHg
- Transducer: 5 uV/VmmHg and 40 uV/VmmHg
- Exciter Voltage: 10 Vac or dc max
- Static Range: 0 mmHg to 250 mmHg, adjustable in 5% increments
- Accuracy: ± 5% or 1 mmHg, whichever is greater
- Waveforms: Static, square wave, and physiological dynamic
Output connector
- Switchcraft 15GM7F
Performance testing
- Static: 0 to 250 adjustable in 5% increments
- Square Wave: 0 mmHg to 250 mmHg adjustable in 5% increments
Respiration
Output signals
- Delta 0.25 Ω to 2.50 Ω Lead I
Impedance: - Base 250 Ω Lead I, 750 Ω Lead II
Impedance: - Rate Range: 0 BPM to 80 BPM
Impedance: - Lead: All leads
Performance testing
- Rate: 0, 10, 20, 40, and 80 per min
- Delta 0.25 Ω, 0.5 Ω, 1 Ω, 1.5 Ω, 2 Ω, and 2.50 Ω
Impedance:- Coincidence check
General (six channels) - ECG/arrhythmia, respiration, arterial pressure, PA pressure, RA pressure, and auxiliary channel for optional cardiac output, CO2, and other parameters.
Fundamental Rhythms and Sequences
Normal sinus rhythm
Sinus tachycardia and bradycardia
Ventricular and sinus asystole
Atrial tachycardia
Atrial flutter and fibrillation
AV blocks (1st degree, 2nd degree Mobitz I and II, and 3rd degree)
Unifocal and multifocal PVCs
Ventricular tachycardia
Ventricular fibrillation
PVCs at 1 to 35/min for any rhythm
Couplet
Triplet
Bigeminy
Junctional
Accelerated junctional
PACs and PJCs
Atrial tachycardia with aberrant conduction
Idioventricular
Agonal
ST-segment elevation and depression
Pacemaker (atrial, ventricular, and AV sequential)
Failure to capture and sense
Bundle-branch block
Cardiac-failure sequence
Conversion sequence
大正医科器械について 資料請求・お問い合わせ
お電話でのお問い合わせはこちら FAXでのお問い合わせはこちら

インターネットお問い合わせフォーム

推奨環境